The portable crono SC 50 SC pumps are designed for subcutaneous infusion of palliative care drugs in terminal patients and for subcutaneous infusion of Treprostinil
The pump is indicated in subcutaneous infusions allowing 4 modes of administration:
Continuous
Bolus on request
Clinical bolus (managed by the doctor)
Combined (continuous + bolus on demand + clinical bolus)
The pump is equipped with a keypad lock to prevent incorrect or unauthorised administration.
The liquid crystal display shows physician and patient information on pump programming, operation and diagnostics.
Pump dimensions | 76 x 49 x 29 mm (3.00 x 1.94 x 1.15 in). |
---|---|
Pump dimensions (with protective wings) | 138 x 49 x 29 mm (5.43 x 1.94 x 1.15 in). |
Pump weight | 115 g (including battery). |
Weight (with protective wings) | 127 g (4.46 oz.), including battery. |
Battery | 3 Volt Lithium battery, mod. 123 A (battery life approximately 100 infusions). |
Single-use reservoirs | Special 10 and 20 ml reservoirs with “Luer-Lock” universal safety fitting. |
Administration doses | Selectable, from 1 ml to 20 ml in 1 ml increments. |
Priming function | Only available at the start of the infusion. The maximum volume that can be delivered is 1500 μl in 5 μl increments. |
Flow rates selectable | From 5 to 2000 μl/h in 1 μl/h increments: It is possible to disable this function by setting 0 μl/h |
Bolus dose | Selectable, from 10 μl to 2000 μl in 10 μl increments. It is possible to disable this function by setting 0 μl. |
Clinical bolus | Selectable, from 10 μl to 2000 μl in 10 μl increments. It is possible to disable this function by setting 0 μl. |
Minimum interval between bolus doses |
|
Bolus doses in one hour | From 0 to 12 bolus doses. The function may be disabled by setting no,Ln. |
Flow rate precision | +/-2%. |
Maximum occlusion pressure | 2,5 bar +/- 1,5 (PL1), 4 bar +/-2 (PL2), 5,5 bar +/- 2 (PL3). |
Display | Liquid crystal display (dimensions 1.1 x 2.8 cm). |
Motor | Coreless DC motor. The microcontroller controls the rotation speed using an infrared encoder. |
Safety circuits | They check that the device is functioning correctly, intervening in the event of any fault with acoustic signals and messages on the display. |
Anti free-flow system | To avoid a non controlled flow inside the delivery set due to the action of the force of gravity. |
Supplied standard equipment | Ambulatory infusion pump, infusion pump carry-case, elastic belt, collar strap, fabric case, 2 batteries, battery cover opening tool, 2 user manuals (doctor and patient). |
Guarantee | Two years for manufacturing defects. |
Pump dimensions | 84 x 55 x 42 mm (3.31 x 2.17 x 1.65 in). |
---|---|
Pump weight | 139 g (including battery). |
Battery | 3 Volt Lithium battery, mod. 123 A (battery life approximately 100 infusions). |
Dedicated single-use reservoirs | With a 50 ml capacity and a “Luer-Lock” universal security attachment. |
Partial volume | Selectable, from 1 to 50 ml in 1 ml increments. |
Priming function | Only available at the start of the infusion. The maximum volume that can be delivered is 1.5 ml. |
Flow rates Selectable | from 0.05 to 5.00 ml/h in 0.01 ml/h increments. It is possible to disable this function by setting 0 ml/h. |
Bolus dose | Selectable from 0.10 to 2 ml in 0.02 ml increments. It is possible to disable this function by setting 0 ml. |
Clinical bolus | Selectable from 0.10 to 2 ml in 0.02 ml increments. It is possible to disable this function by setting 0 ml. |
Minimum interval between bolus doses |
|
Bolus doses in one hour | From 0 to 12 bolus doses. It is possible to disable this function by setting no,Ln. |
Flusso precision | +/-3%. |
Maximum occlusion pressure | 1,8 bar +/-0,8 (PL1) - 2,2 bar +/-1 (PL2) - 3,2 bar +/-1,2 (PL3). |
Setting memory | The settings are automatically entered in a flash memory which is retained (with the exception of the partial syringe volume) even if the battery is removed from the device. |
Display | Liquid crystal display (0.43 x 1.1 in, 1.1 x 2.8 cm). |
Motor | Coreless DC motor. The microcontroller controls the rotation speed using an infrared encoder. |
Safety circuits | They check that the device is functioning correctly, intervening in the event of any anomaly with acoustic signals and messages on the display. |
Anti-free-flow safety system | To avoid a non controlled flow inside the delivery set due to the action of the force of gravity. |
Supplied standard equipment | Ambulatory infusion pump, infusion pump carry-case, elastic belt, collar strap, fabric case, 2 batteries, battery cover opening tool, 2 user manuals (doctor and patient). |